11.5 4-steps check mode 1
In 4-steps check mode 1, up to four comparison values and weighed values can be
compared the results output.
Set function SQF-01 to [2] to use this mode.
The setting of each brand number for 4-steps check mode can be set by F-20.
Setting brand comparison targets (Targeted functions from F21 to F26)
Note:
Output for each comparison is synchronized with A/D sampling.
11.5.1
S1, S2, S3 and S4 comparison operation
Operating conditions for check modes for S1, S2, S3 and S4 can be selected from [Set
value or more] and [Set value or less].
Select in function mode (F-22).
The default for S1, S2, S3 and S4 is [Set value or more].
Note:
-
The ON/OFF status for each output may difer depending on selection. If an
inappropriate mode is selected, the ON/OFF status may be incorrect, leading to
malfunction of peripheral instruments.
-
The set points of S1, S2, S3, and S4 are shared between 4-steps check mode 1 and
4-steps check mode 2.
Operation at the comparison value for every comparison target and output operation in
the case of S1 are shown below.
The same applies to S2, S3 and S4.
-
[Set value or more] is selected.
S1 output turns ON at S1 set value ≤ comparison value.
Comparison
target value (larger)
Comparison
target value (smaller)
S1 Output ON
S1 Output OFF
S1 setting value
-
[Set value or more] is selected.
S1 output turns ON at S1 set value ≥ comparison value.
11 Weighing mode
Weight indicator CSD-903
Minebea Intec
EN-105