Item
Function No.
Number Return command set value
Overload display condition
71
03
0: Greater than ?maximum load + 9 D?
1: When the amount exceeds maximum load
±110%
2: Less than –20 D and larger than maximum
load +9 D
Unit
71
05
0: No unit; 1: g; 2: kg; 3: t; 4: lb; 5: N; 6: kN
Bridge power supply
71
07
0: 10 V DC; 1: 5 V DC, 2: 2.5 V DC
Zero set operating condition
71
10
0: Operation when stable
1: Unconditional operation
Zero set efective range
71
11
00 to 30 (Unit : ±1% of maximum load)
Zero tracking target
71
12
0: Gross weight and net weight (load display)
1: Gross weight (gross weight display only)
ZERO tracking data width
71
13
00 to 99; Unit: 0.1 D)
Zero tracking time width
71
14
00 to 99; Unit: 0.1 s)
Power on zero operation
71
15
0: Invalid; 1: Efective
Tare weight cancellation opera-
ting condition
71
16
0: Operation when stable and 0 < GROSS
WEIGHT
≤ MAXIMUM LOAD
1: Unconditional operation and 0 < GROSS
WEIGHT
≤ MAXIMUM LOAD
2: Operation when stable and 0 < GROSS WEIGHT
≤| MAXIMUM LOAD|
3: Unconditional operation and 0 < GROSS
WEIGHT
≤| MAXIMUM LOAD|
Date saved
71
17
0: RAM; 1: EEPROM
Power on clear
71
18
000 to 111 (0: Invalid; 1: Efective)
10
1
: Tare weight cancellation data
10
2
: Preset tare weight cancellation data
10
3
: Zero set and zero tracking data
Setting method for gravity acce-
leration compensation
71
25
0: Set the district number
1: Set the numerical value for gravity acceleration
District number for place of use 71
26
01 to 16 (Unit: District)
District number for place of cali-
bration
71
27
01 to 16 (Unit: District)
Gravity acceleration setting for
place of use
71
28
9.000 to 9.999 (Unit: m/s
2
)
Gravity acceleration setting for
place of calibration
71
29
9.000 to 9.999 (Unit: m/s
2
)
18 Options