Ensure that nothing is on the load cell other than the tare weight outside maximum
load.
[TZERO]
Stand-by condition
Error display in zero point calibration
[TE-L]: The display blinks for about two seconds if load cell output is –2.5
mV/V or less (refer to Chapter
[TE-H]: The display blinks for about two seconds if load cell output is 2.5
mV/V or more (refer to Chapter
).
6.7 CAL-switch
Lock calibration settings to protect them from changes. Use the CAL switch to lock the
calibration settings.
1. Switch the display to [TZERO] by pressing the [F] key on the [TARE WEIGHT]display.
2. The display of [TZERO] blinks after pressing the [F] key and zero point calibration
starts.
3. Press the [F] key when the stability mark light is on.
[END]
Zero point calibration is complete.
4. Press the [F] key to quit calibration mode for the zero point only.
[CCAL]
[CCAL] is displayed. The set data is saved to the internal memory.
5. Set to stand-by condition by pressing the [ON-OFF] key.
Calibration of the zero point only is complete.