7.2 C function data functions
Item
Function No.
Setting value Contents
0*
No decimal point
1
1234.5
2
123.45
3
12.345
Decimal point display position
CF-01
4
1.2345
0
20 times/s
A/D sampling rate
CF-02
1*
200 times/s
0
Greater than |maximum load +9D|
1
When the amount exceeds maximum load
±110%
Overload display condition
CF-03
2*
Less than –20D and larger than maximum
load +9D
0*
No unit
1
g
2
kg
3
t
4
lb
5
N
Unit
CF-05
6
kN
0*
10 V DC
1
5 V DC
Excitation
CF-07
2
2.5 V DC
0*
Net weight sign not inverted
Net weight sign inversion
CF-08
1
Net weight sign inverted
0
Operation when stable
Zero set operating condition
CF-10
1*
Unconditional operation
00 ~ 30
Zero set efective range
CF-11
02*
Unit: ±1% of maximum load
0
Gross weight and net weight (load display)
Zero tracking target
CF-12
1*
Gross weight (gross value display only)
00 ~ 99
Zero tracking data width
CF-13
01*
Unit: 0.1; D
00: Zero tracking OFF
00 ~ 99
Zero tracking time width
CF-14
00*
Unit: 0.1 s
00: Zero tracking OFF