The default is [Operation when stable and 0 < gross weight ≤ maximum load].
Note:
-
Tare weight cancellation is not accepted when the gross weight value exceeds the
irst range.
-
Preset tare weight cancellation is cleared and tare weight cancellation is executed if
preset tare weight cancellation is executed. (Tare weight cancellation cannot be
used together with preset tare weight cancellation.)
-
When the zero set is executed with gross weight value display, tare weight
cancellation is cleared.
8.11 Clear at power on
Setting clear at power on is selected by C function CF-18.
Clear at power on clears [Data of tare weight cancellation], [Data of preset tare weight
cancellation] and [Data of zero set and zero tracking] when the power is turned on or
when the display is turning on.
Setting: 000 to 111; 0 Invalid; 1: efective; Default: 000
10
0
: Tare weight cancellation data
10
1
: Preset tare weight cancellation data
10
2
: Zero set and zero tracking data
8.12 Setting in set mode 2 or preset tare weight set
Proceed to set the unit display in the sub-display section and the stability detection in the
condition display in set mode 2 or preset tare weight set.
8.12.1
Stability detection
Selection of whether stability detection is efective or invalid is made by C function CF-20.
The default is [Invalid].
8.12.2
Unit display in the sub-display section
Selection of whether the unit display in the sub-display section is efective or invalid is
made by C function CF-21.
The default is [Invalid].
8.13 Gravity acceleration compensation
A SPAN error arises due to diference in gravity acceleration if the places of calibration
and usage difer.
The gravity acceleration compensation function compensates for SPAN error by setting
diferent gravity accelerations for the places of calibration and usage. When the
calibration place and the using place are the same, it is not necessary to compensate for
gravity acceleration.
8.13.1
Setting method for gravity acceleration compensation value
The setting method for gravity acceleration compensation value is selected by C function
CF
‑
25.
Weight indicator CSD-903
8 Various operations by C function data
EN-60
Minebea Intec
Summary of Contents for CSD-903
Page 243: ......