BCD output check
↓
Correct connecting?
→ No →
↓
Yes
↓
Connect wires surely with an exter-
nal equipment, refer to Chap-
ter
Except for the BCD are outputted.
→ Yes →
↓
No
↓
Match the setting concerning the
BCD output to the speciication of
external equipment connected, re-
fer to Chapter
The intended value is not output-
ted.
→ Yes →
↓
No
↓
Select the target of BCD output, re-
fer to Chapter
The BCD output does not change.
→ Yes →
Make the HOLD input opened.
↓
No
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.
RS-232C check
↓
Correct connecting?
→ No →
↓
Yes
↓
Connect wires surely with an exter-
nal equipment, refer to Chap-
ter
The Communication protocol is
correctly. set.
→ Yes →
↓
No
↓
Set it according to the host's speci-
ications, refer to Chapter
.
Command/data is set suitably.
→ Yes →
↓
No
↓
Use the correct command/data, re-
fer to Chapter
.
The host speciies the low control.
→ Yes →
↓
No
↓
Set the low control from [YES] to
[NONE].
In Stream mode, intended value is
not sent.
→ Yes →
Select the output target in the
Stream mode.
↓
No
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.
20 Troubleshooting
Weight indicator CSD-903
Minebea Intec
EN-227