9.2 Details of function data
Item
Function No.
Setting
Contents
Digital ilter setting
F-01
001 ~ 256
016
Unit: Moving average time 1 time)
0
2 Hz
1*
4 Hz
2
6 Hz
3
8 Hz
Analog ilter setting
F-02
4
10 Hz
0*
4 times/s
Display times
F-04
1
20 times/s
Stabilization ilter setting F-05
001 to 256
092*
Unit: Moving average time 1 time)
Stabilization ilter data
width
F-06
000 to 999
020*
Unit: 0.1 D
000: Stabilization ilter OFF
Stabilization ilter time
width
F-07
00 to 99
05*
Unit: 0.1 s
00: Stabilization ilter OFF
Lock key 1
F-08
0000 to 1111
0000*
0: Invalid
1: Valid
10
0
: ON/OFF key
10
1
: Setting key
10
2
: Accumulation/◀ key
10
3
: Tare weight set/
►
key
Lock key 2
F-09
0000 to 1111
0000*
0: Invalid
1: Valid
10
0
: Tare weight cancellation/▲ key
10
1
: Net or gross weight/▼ key
10
2
: Zero set key
10
3
: F/
↵
key
Stability detection data
width
F-10
000 to 999
020*
Unit: 0.1 D
00: Stabilization ilter OFF
Stability detection time
width
F-11
00 to 99
10*
Unit: 0.1 s
00: Stabilization ilter OFF
0*
Invalid
1
Efective
Preset tare weight cancel-
lation operating condition
F-15
2
Net weight ofset operation
9 Function mode
Weight indicator CSD-903
Minebea Intec
EN-67
Summary of Contents for CSD-903
Page 243: ......