8.8 ZERO tracking
The zero tracking function compensates the gradual zero drift under constant conditions
and stabilizes the zero point.
8.8.1
ZERO tracking target
The ZERO tracking target is set by C function CF-12.
[Gross weight value or net weight value (for weight display)] or [Gross weight value (only
with gross value display)] can be selected.
The default is [Gross weight value].
8.8.2
ZERO tracking data width
The ZERO tracking data width is set by C function CF-13.
Setting range: 00-99; Unit: 0.1 D; 00: OFF
The default is [00].
8.8.3
ZERO tracking duration
The ZERO tracking target is set by C function CF-14.
Setting range: 00-99; Unit: 0.1 s; 00: OFF
The default is [00].
Example:
The data width over which ZERO tracking is executed is set by C function CF-13.
The ZERO tracking data width per set value [n] is obtained in the display conversion by
the following formula.
ZERO tracking data width = Set value n for CF-13× 0.5 × Scale interval
When the set value for C function CF-13 = 10 and the scale interval D = 5, then ZERO
tracking data width = 10 × 0.5 × 5 D = 25 D.
Setting of CF-14
Less than setting of CF-14
Setting of CF-14
Setting of CF-14
+ Data width of
ZERO tracking
– Data width of
ZERO tracking
ZERO
Display
Weight indicator CSD-903
8 Various operations by C function data
EN-58
Minebea Intec
Summary of Contents for CSD-903