6.3.7.1
Method by weight
Place the weight corresponding to the mass on the load cell (Setting described in
Chapter
[SPAN]
Under the SPAN calibration
[SPAN]
Error display in SPAN calibration:
[SP-L]: The display blinks for about 2 seconds if ([Load cell output voltage
at span point] – [Load cell output voltage at zero point]) ≤ 0.0 mV/V (refer to
Chapter
[SP-H]: The display blinks for about two seconds if load cell output is 3.1
mV/V or more (refer to Chapter
).
6.3.7.2
Method by numeric output
Calibration by inputting load cell output voltage.
[SPAN]
[Z MV]
[
✶
.
✶✶✶✶✶
] is the saved load cell output voltage.
Set the value in mV/V to the diference between the measurement load cell output
voltage zero point and the output voltage corresponding to the maximum load mass.
Saved load cell output voltage:
mV/V
[TARE WEIGHT] [NET-GROSS WEIGHT]: The changed digit value adjusts.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[ZERO]: The displayed value changes to zero.
[SET]: Interrupts the setting and returns to the CCAL display
[F]: The displayed value is saved. Proceed to the next step.
1. Press the[F] key with the weight on the load cell.
2. Press the [F] key when the[SPAN] display blinks and the STABLE mark is lit.
[END]
▼ End.
[SPAN] is saved, and [END] is displayed.
1. When the[PRESET TARE WEIGHT] key is pressed, [Z MV] is displayed.
2. When the[F] key is pressed, [
✶
.
✶✶✶✶✶
] is displayed.
Weight indicator CSD-903
6 Calibration
EN-42
Minebea Intec