13 Storage location of setting data
In this unit, each data is recorded in RAM and EEPROM as follows:
The data in EEPROM can be stored almost permanent due to nonvolatile.
Moreover, RAM is backed up with the battery. The backup time is about ten years in the
room temperature.
13.1 Data stored in internal RAM
-
Tare weight cancellation data
-
ZERO set data
-
ZERO tracking data
-
Preset tare weight cancellation data
-
Accumulated value, Accumulated count data
13.2 Data stored in EEPROM
-
Calibration data
-
Function data
-
C function data
-
SQ function data
-
Digital linearize data
13.3 Change the storage location for data
The storage target is selected by C function CF-17
The storage target can be selected from [RAM] and[EEPROM].
Storage locations are as follows.
[RAM] is the default.
-
Weighing comparison value 1 setting data
-
Weighing comparison value 2 setting data
Note:
-
Set the storage destination to Internal RAM when the set value is continuously
changed and the number of EEPROM rewrites might exceed 1 000 000.
-
Do not change in the setting during weighing because the A/D sampling rate
decreases temporarily while rewriting the set value.
Set the storage destination to Internal RAM to facilitate changes setting values
during weighing.
13 Storage location of setting data
Weight indicator CSD-903
Minebea Intec
EN-123