6.5 Digital linearization
After calibration, a weighing error of several scale intervals may arise between ZERO and
SPAN (maximum load) due to the inluence of the load cell.
Digital linearization is a function that compensates up to three points, excluding ZERO
and SPAN, to reduce weighing error.
Weight value
Error
SPAN
ZERO
Weight value
0
0
ZERO
SPAN
Correction point 3
Correction point 2
Correction point 1
After linearization
Before linearization
Display
Display
6.5.1
Switch to digital linearization mode
Weighing mode
Stand-by condition
[CFUNC]
[CAL]
[CCAL]
[CADJ]
[LNR]
▼ Digital linearization mode
6.5.2
Setting digital linearization
[LNR]
[POINT]
1. Set the condition from normal weighing mode to stand-by condition by pressing the
[ON-OFF] key.
2. Press the [ON-OFF] key while pressing the [SET] key. [CFUNC] is displayed.
3. Press the [F] key after pressing the [NET-GROSS] key. [CCAL] is displayed.
4. [LNR] is displayed after pressing the [NET-GROSS] key twice to enter digital
linearization mode.
Weight indicator CSD-903
6 Calibration
EN-46
Minebea Intec
Summary of Contents for CSD-903
Page 243: ......