External control I/O check.
↓
Connecting wiring are cor-
rected?
→ No →
↓
Yes
↓
Connecting surely, refer to Chapter
The external control I/O is
set?
→ No →
Set the external control I/O.
↓
Yes
↓
External control input is
wrong.
→ Yes →
↓
No
↓
Inform Minebea Intec about the contens of
failure and situation at site in details.
50 ms or more of input sig-
nal is set for the pulse input.
→ No →
↓
Yes
↓
Set the pulse input value 50 ms or more.
50 ms or more of input sig-
nal for the item of Level in-
put.
→ No →
↓
Yes
↓
Set the pulse input value 50 ms or more.
Inform Minebea Intec about the
contens of failure and situation
at site in details.
20 Troubleshooting
Weight indicator CSD-903
Minebea Intec
EN-225