②
↓
1.
Remove the connection cable for the load cell from the terminal block for load cells.
2.
Measure the voltage between A-C after short-circuiting A-F and C-G of terminal
block.
Set the load cell tester to DC voltage measurement.
↓
The voltage between A and C is
stable in the set value of excitation
voltage.
→ No →
Inform Minebea Intec about the
contens of failure and situation at
site in details.
↓
Yes
↓
1.
Shorten the B, D and C in the connector.
2.
Conirm the output voltage of the load cell, refer to Chapter
.
↓
The display stabilizes with the opti-
onal value.
→ No →
↓
Yes
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.
1.
Check the wiring status, refer to Chapter
.
2.
Check whether interference sources are present, e.g. converter.
↓
↓
Same condition
→ No →
Start weighing
↓
Yes
↓
Inform Minebea Intec about the
contens of failure and situation at
site in details.
20 Troubleshooting
Weight indicator CSD-903
Minebea Intec
EN-221