15 2-pin method serial interface (S-I/F)
15.1 Interface speciication
Speciication
Contents
Baud rate
600 bps
Data length
8 bit
Parity bit
Odd number
Stop bit
1 bit
Start bit
1 bit
Sending date
Binary code and BCD
15.2 Data format
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
F1
F2
F3
LEN FUN1 FUN2 ST1
ST2 OVR ERR
G1
G2
G3
N1
N2
N3
T1
T2
T3
BCC
1.
2.
F1 to F3
3.
LEN
4.
FUN1
5.
FUN2
6.
ST1
7.
ST2
8.
OVR
9.
ERR
10.
G1 to G3
11.
N1 to N3
12.
T1 to T3
13.
BCC
1) FUN1
7
6
5
4
3
2
1
0
2) ST1
7
6
5
4
3
2
1
0
3) ST2
7
6
5
4
3
2
1
0
G1/N1/T1
G2/N2/T2
G3/N3/T3
DP
10
4
10
3
10
2
10
1
10
0
Load data and BCD data
7
6
4
5
7
6
0
0
0
1
10
1
1
0
10
2
1
1
10
3
INTERVAL
INTERVAL
Space for 15 bit(25 ms) or more (MARK SIGNAL)
0FFH code
11H code
Print status
00H code
Condition
Condition
00H code
Condition
GROSS data
NET data
TARE data
Block Checking Character from LEN to T3
Printing status
On [1] means NET display
On [1] means GROSS display
On [1] means preset tare is valid
4) GROSS data / NET data / TARE data
On [1] means minus
Decimal point
Decimal
point
None
15 2-pin method serial interface (S-I/F)
Weight indicator CSD-903
Minebea Intec
EN-131
Summary of Contents for CSD-903