140 & 180 ANALOG MODELS
PAGE 1
Issue Rev C 1221
Copyright Quincy Lab Inc. 2021
General Purpose Incubators
OPERATING MANUAL
Model Series 140 & 180
HEAT/CYCLE
POWER
Model 12-140 Incub
ator
Interior Dimensions
INCHES W x H x D
(CM) W x H x D
Exterior Dimensions
INCHES W x H x D
(CM) W x H x D
12x10x10
18x16x12
18x16x12
12x10x10
13x15x11
19x21x13
13x15x11
19x21x11
Weight (lbs)
19 lbs
33 lbs
19 lbs
33 lbs
Standard Electrical
31x25x25
46x41x30
46x41x30
31x25x25
33x38x28
48x53x33
33x38x28
48x53x33
SPECIFICATIONS
Temperature Range
A 2oC to 62oC
Common Unit Construction
Exterior
: Powder-Coated Steel
Interior:
Aluminum
Insulation:
Fiberglass
Door:
140: Acrylic, 180: Steel Insulated
Thermo-control:
Bi-Metal
Heater:
Resistive-Tubular Incoloy
* Standard models voltage only, optional 230 voltage available. Check label on back of unit.
VOLTS / WATTS
115 / 120*
115 / 235*
115 / 270*
115 / 385*
HEAT/CYCLE
POWER
Model 10-180 Incubat
or
A 3oC to 94oC
Cubic Foot Capacity
.7 ft
3
2.0 ft
3
.7 ft
3
2.0 ft
3
Model 10-180
Model 12-140
Common Unit Specifications
Operating Environment
:
Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Pollution Degree 2, ambient temperature 10
o
C/50
o
F to 35
o
C/95
o
F,
80% RH maximum.
Storage Temperature:
-10
o
C/14
o
F to 70
o
C/158
o
F, 70% RH maximum.
Approvals:
Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
Compliance:
UL Standard 61010-1, IEC 61010-1, 2nd Edition.
MODEL
10-140
MODEL
12-140
MODEL
10-180
MODEL
12-180
uincy
ab