[SQF-02]
Select another SQ function number and repeat this procedure.
[SQFUNC]
Press the [SET] key to switch to weighing mode.
▼ Weighing mode
Do not change in the setting during weighing because the A/D sampling rate decreases
temporarily while rewriting the set value.
11.2 Setting mode
Set the value at which each weighing operation is executed in setting mode.
Setting vary with selected weighing mode.
11.2.1
Weighing comparison value 1 setting mode
Set the weighing comparison value for simple mode and sequential mode.
11.2.1.1
Setting method for weighing comparison value 1 setting mode
Weighing mode
[FUNC.]
[COMP]
[SET1],
[SET2]
The display corresponding to SET1 to SET8 and B0 to B7 blinks.
[TARE WEIGHT] [NET-GROSS WEIGHT]: Adjust the value of the digit to be changed.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[SET]: The display returns to [COMP].
[F]: The displayed value is saved. Proceed to the next step.
The display corresponding to SET1 to SET8 blinks.
5. Press the [F] key. The setting is registered. The following registered SQ function
number is displayed.
6. Press the [SET] key after registration is complete to switch to the [SQFUNC] display.
1. Change from normal weighing mode to the [FUNC] display by the [SET] key.
2. [COMP] is displayed after pressing the key [NET-GROSS WEIGHT] key twice to switch
to weighing comparison value 1 setting mode.
3. Press the [F] key[to display SET1].
4. Press the [F] key. Set the selected comparison value.
Weight indicator CSD-903
11 Weighing mode
EN-90
Minebea Intec