7 C function mode
The various calibration functions become efective by setting the C function data.
7.1
C function mode setting method
Weighing mode
Stand-by condition
[CFUNC]
[CF-01],
[CF-03]
Select C function number to be changed.
[TARE WEIGHT] [NET-GROSS WEIGHT]: Adjust the value of the digit to be changed.
[ACCUM] [PRESET TARE WEIGHT]: Select the changed digit.
[ZERO]: Change the display value to [0].
[SET]: Return to the [CFUNC] display.
[F]: Saves the displayed value and proceeds to the next step.
[ON-OFF]: Quit C function mode without changing other settings.
,
[CFUNC]
Stand-by condition
1. Set the condition from normal weighing mode to stand-by condition by pressing the
[ON-OFF] key.
2. Press the[ON-OFF] key with the[SET] key. [CFUNC] is displayed.
3. [CF-01] is displayed by pressing the [F] key.
4. Press the [F] key to change the selected C function mode setting.
5. Press the [F] key.
[CF-05]
Registration is complete.
Repeat this process if another C function number setting is to be changed.
The changed content is registered. The following changed C function number is
displayed.
6. Press the [SET] key after registration is complete. [CFUNC] is displayed.
7. Set to stand-by condition by pressing the [ON-OFF] key.