8.14.3
Setting the limit value for the second range
The limit value for the second range is set by C function CF-42.
Setting range: 0 to 999999
The default is [003000].
8.14.4 Setting the scale interval of the third range
The scale interval of the range is set by C function CF-43.
The scale interval can be selected from [1 scale interval], [2 scale interval], [5 scale
interval], [10 scale interval], [20 scale interval] and [50 scale interval].
The default is [10 scale interval].
8.14.5 Setting the limit value for the third range
The limit value for the third range is set by C function CF-44.
Setting range: 0 to 999999
The default is [006000].
Example 1
Gross weight value
First range: 300.0 g (limit for the second range), scale interval 0.1 g (calibration scale
interval)
Second range: 600.0 g (limit for the third range), scale interval 0.5 g (second range scale
interval)
Third range: 1 000.0 g (maximum load), scale interval 1 g (third range scale interval)
0 g
300 g
600 g
1000 g
Scale interval 0.1 g
Scale interval 0.5 g
Scale interval 1 g
Weighing value
The weighing values from 0 to 300 g becomes the irst range (scale interval 0.1 g).
The weighing values from 300 to 600 g becomes the second range (scale interval 0.5 g).
The weighing values from 600 to 1,000 g becomes the second range (scale interval 1 g).
8 Various operations by C function data
Weight indicator CSD-903
Minebea Intec
EN-63