Value of the load cell weight
[ZERO]
▼ Zero point calibration
6.3.6
Zero point calibration
[ZERO] is displayed when SPAN calibration starts.
Select the zero calibration method
-
Method by weight data (status of initial load)
⇨
Press the [F] key and execute the
"Method by measuring data".
Execute zero calibration with nothing on the load cell.
-
Method by load cell numeric output
⇨
Press the [PRESET TARE WEIGHT] key and
execute the "Method by numeric input".
Execute zero calibration by inputting the output voltage value of the load cell when
nothing is on the load cell.
6.3.6.1
Method by weight data
Method by weight data (status of initial load).
Ensure that the load cell is in an unladen state.
[ZERO]
Under the zero calibration
[ZERO]
Error display in zero point calibration
[TE-L]: The display blinks for about two seconds if load cell output is –2.5
mV/V or less (refer to Chapter
[TE-H]: The display blinks for about two seconds if load cell output is 2.5
mV/V or more (refer to Chapter
).
6.3.6.2
Method by numeric output
Calibration by inputting load cell output voltage.
[ZERO]
2. Press the [F] key after setting
1. Press the [F] key, with nothing on the load cell
2. Press the [F] key when the[ZERO] display blinks and the STABLE mark is lit.
[SPAN]
▼ SPAN point calibration
The zero point is saved, and [SPAN] is displayed.
Weight indicator CSD-903
6 Calibration
EN-40
Minebea Intec