The default is [05].
Example
The data width over which the stabilization ilter is executed is selected by function F-06.
The stabilization ilter data width for set value "n" is obtained as follows.
Stabilization ilter data width = Set value of F-06 × scale interval
Setting of function F-06 is [10] and the scale interval is [D = 5].
Stabilization ilter data width = 10 × 5 × 0.1 = 5 D
Therefore, the stabilization ilter set by function F-05 is active when the load display
luctuation is within the value set by function F-06 and this condition exceeds the time set
by function F-07.
Time width of
stabilization filter
Data width of Stabilization filter
ON
OFF
Time width of
stabilization filter
Time width of
stabilization filter
Data width of
Stabilization filter
Data width of
Stabilization filter
Digital filter for
stabilization filter
Change width of weight
Change width of load
Change width of weight
Time
Display
Note:
When the digital ilter is set by function F-01, the moving average logic executes
"moving average of stabilization ilter (F-05)" after executing "moving average of digital
ilter (F-01)."
10.5 Key lock function
Key lock function restricts operation to prevent operational error for various front panel
keys.
10.5.1
Key lock 1
Function F-08 sets whether key operation of [ON/OFF], [SET], [ACCUM./◀] and [PRESET
TARE WEIGHT/▶] are[valid] or [invalid].
Setting range: 0000 to 1111; 0: Invalid; 1: Valid
10
0
digit: ON/OFF
10
1
digit: SET
10
2
digit: ACCUM./◀
10
3
digit: PRESET TARE WEIGHT/▶
10 Various operations by function data
Weight indicator CSD-903
Minebea Intec
EN-77