①
Switch to communication calibration mode (refer to Chapter
)
②
Set the scale interval (refer to Chapter
)
③
(3) Set weighing operation (refer to Chapter
)
④
Set the mass for the weight (refer to Chapter
Method by numeric output of load
cell output voltage
Method by weighing operation
⑤
Zero stability check (refer to Chapter
⑦
Numeric input of load cell output voltage
Zero registration (refer to Chapter
)
⑥
Zero registration by weighing operation (refer to
Chapter
⑦
Numeric input of load cell output voltage
SPAN registration (refer to Chapter
)
⑤
SPAN stability check (refer to Chapter
)
⑥
SPAN registration by weighing operation (refer to
Chapter
)
⑪
Calibration registration (refer to Chapter
Note:
-
CSD-903 displays
(-RS-) during communication calibration.
-
When calibration is force inished, all data returns to its condition prior to calibration
without saving.
-
In the case calibration by weighing value registration, register after conirming
stability.
Weight indicator CSD-903
18 Options
EN-206
Minebea Intec