Note:
-
Only during the measurement mode, the output of serial data is done.
-
The output of the data of S-I/F stops when the maximum display is set to six digits.
Set the maximum display within ive digits when you use S-I/F.
-
The output of the data of S-I/F stops when decimal point is from digits. Set the
decimal point less than three digits when you use S-I/F
-
Refer to Chapter
15.3 Explanation of format data
-
Automatic print bit [FUN1: bit 0]
This bit can be used as a trigger at receiving side of the instrument with 2-pin method
serial interface.
-
Display bit [ST1:bit 7, bit 6]
Bit 7: Put [1] when the display of the instrument is gross weight.
Bit 6: Put [1] when the display of the instrument is net weight.
15.4 Explanation of output type
-
Stream mode
Keep outputting the data of gross weight and the net weight in constant intervals.
-
Print signal operation
Print signal operation is selected by function F-37 and can be selected from [OFF],
[Synchronizing with print signal], [Synchronizing with accumulation signal], and
[Synchronizing with inish signal].
When [Synchronizing with print signal] is selected, send with [1] set to the automatic
print bit on the FUN1 of output format synchronized with the [F] key, the external
control input, or automatic printing.
When [Synchronizing with accumulation signal] is selected, send with [1] set to the
automatic print bit on the FUN1 of output format when [Accumulation command] is
input by the function F-50 setting, key, or external control input.
When [Synchronizing with inish signal] is selected, send with [1] set to the automatic
print bit on the FUN1 of output format when [Forced batch inish] is input by the [F]
key or external control input or synchronized with the batch inish signal.
Default is set to [OFF].
Note:
The automatic print bit is usually 0.