Item
Function No.
Number Return command set value
Range switch operation
71
40
0: Range change OFF;
1: Can be changed between two ranges;
2: Can be changed between three ranges
Setting the scale interval of the
second range
71
41
0: 1 scale interval; 1: 2 scale interval;
2: 5 scale interval; 3: 10 scale interval;
4: 20 scale interval; 5: 50 scale interval
Setting the boundary value for
the second range
71
42
000000 to 999999
Setting the scale interval of the
third range
71
43
0: 1 scale interval; 1: 2 scale interval;
2: 5 scale interval; 3: 10 scale interval;
4: 20 scale interval; 5: 50 scale interval
Setting the boundary value for
the third range
71
44
000000 to 999999
Analog output target
71
70
0: Display interlock; 1: Gross value;
2: Net weight
Display value at analog output
of 4 mA DC (0 V DC)
71
71
–999999 to 999999
Display value at analog output
of 20 mA DC (10 V DC)
71
72
–999999 to 999999
Calibration data for scale
interval (Reference)
71
90
0: 1 scale interval; 1: 2 scale interval;
2: 5 scale interval; 3: 10 scale interval;
4: 20 scale interval; 5: 50 scale interval
Calibration data for maximum
load (Reference)
71
91
000000 to 999999
Calibration data for weight (Re-
ference)
71
92
000000 to 999999
Calibration data for zero
mV/V (Reference)
71
93
–3.1000 to 3.1000 (Unit: 0.00001 mV/V)
Calibration data for span
mV/V (Reference)
71
94
–0.00001 to 3.1000 (Unit: 0.00001 mV/V)
Stability detection time width at
calibration
71
97
00 to 99; Unit: 0.1 s)
Weighing mode
72
01
0: Simple mode
1: Sequential mode
2: 4-step check mode 1
3: 4-step check mode 2
Control mode
72
02
0: Batching mode; 1: Discharge mode; 2:
2: External input changeover
Comparison signal operation
72
05
0: ANY TIME; 1: IN STABLE MODE;
2: SYNC. WITH INFLOW FIN;
3: INFLOW FIN + HOLD
Weight indicator CSD-903