14: D. (Fine) FLOW/S3
Lights when ine low output signal is [ON]in sequential
mode Lights when S3 output signal is [ON]in check mo-
de 1 or 2
15: OVER/S4
Lights when the judgment condition is over in sequential
mode Lights when S4 output signal is [ON]in check mo-
de 1 or 2
16: UNDER
Lights when the judgment condition is under in sequen-
tial mode
17: FULL
Lights when the judgment condition is full in sequential
mode
18: During weighing
Output during weighing
19: INFLOW FINISH
Output on inishing weighing
20: DISCHARGE (GATE OPEN) Output during discharge
21: DISCHARGE FINISH
Output on inishing discharge
22: RUN
Output in weighing mode
The default is [OFF].
10.15 Memory clear
Memory clear is executed by function F-99.
The function setting returns to the default setting.
[CF-99]
[F.CLR]
[F.CLR]
[F-01]
Note:
Do not clear the memory during weighing as it can result in malfunction.
1. If the [F] key is pressed with[CF-99] displayed, [LNR.CLR] blinks.
2. Press the [ON-OFF] key to discontinue memory clear.
Stand-by condition
The display returns to stand-by mode, and digital linearization is not clear.is not
executed.
3. When [F.CLR] is blinking, press the [F] key to turn of the display and switch to [F-01]
display. Memory clear is complete.
10 Various operations by function data