6.3 Calibration procedures
6.3.1
Calibration low
Step 1
Connect to the load cell
Connect the load cell with the instrument.
Step 2
Turn on the power supply.
Power up the system.
Step 3
Set the bridge power supply.
Refer to Chapter
.
Set the bridge power supply.
Step 4
Leave for 10 min. after the po-
wer is turned on.
Energize the unit for about 10 min. to stabilize the unit and load
cell.
Step 5
Switch to calibration mode.
Refer to Chapter
Switch to calibration mode.
Step 6
Set the scale interval.
Refer to Chapter
Set the scale interval.
Step 7
Set the maximum load.
Refer to Chapter
Set the maximum load.
Step 8
Set the mass for the weight.
Refer to Chapter
.
Set the mass of the weight on the load cell.
Step 9
Zero point calibration
Refer to Chapter
.
Execute zero point calibration with nothing on the load cell or by
setting the output voltage of the load cell zero point.
Step 10 SPAN calibration
Refer to Chapter
.
Execute SPAN calibration with a weight on the load cell or by set-
ting the output voltage of the load cell SPAN point.
Step 11 Quit calibration.
Refer to Chapter
Exit calibration mode.
Carry out settings from step 12 as follows as necessary
Step 12 Set the decimal point display position
Refer to Chapter
Step 13 Set the units
Refer to Chapter
Step 14 Digital linearization
Refer to Chapter
.
Step 15 Automatic range switch
Refer to Chapter
.
Step 16 Gravity acceleration
Refer to Chapter
6 Calibration
Weight indicator CSD-903
Minebea Intec
EN-37