3
Synchronized with print signal
Data is output synchronized with the print signal (F key, external control input or
automatic print).
4 Synchronized with accumulation signal
Data is output when [Accumulation command] is input by function F-50 setting, key
or external control input.
5 Synchronized with inish signal
Data is output when [Forced inlow inish] is input by the [ACCUM] key or external
control input, or synchronized
with the inlow inish signal.
Note:
the RS-485 interface continues to output data at a ixed rate when stream mode is
selected.
Control cannot be carried out by the host.
16.3.2
Speciications of communication by RS-485
RS-485 communication speciications are selected by function F-42.
Function No.
Set value
Details
F-42
00000 to 13121
0
0 0
0 0
0 : 7 bit, 1 : 8 bit
0 : 1 bit, 1 : 2 bit
0 : 1 200 bps,
1 : 2,400 bps
2 : 4 800 bps,
3 : 9 600 bps
4 : 19 200 bps
5 : 38 400 bps
0 : CR, 1 : CR + LF
Data bit length
Parity bit
Stop bit
Baud rate
Terminator
0 : No parity
1 : Even parity
2 : Odd parity
The default is [13020]
16.3.3
RS-485 address setting
The address of the RS-422/485 interface is set by function F-43.
Setting range: 00 to 31
The default is [00].
16.3.4
Delay time of RS-485 return data
Return data from the CSD-903 side can be delayed when the transmitting terminal on the
host side enters a state of low impedance.
Delay time for return data from the RS-485 interface is set by function F-45.
Setting range: 00 to 99; Unit: 10 ms
The default is [01].
Weight indicator CSD-903
16 Standard RS-485 Communication
EN-134
Minebea Intec