F-5
SIMV-VC and SIMV-PC Modes
Tinsp (s)
1.5
SIMV Rate (BPM)
10
Psupp (cmH2O)
15 cmH
2
O
Trigger Window
25%
Trigger Level
-2 cmH
2
O (pressure-triggered)
3.0 L/min (L/min) (flow -triggered)
PSV Insp Termination Level
25%
Ventilator alarm limits
FiO2 High Limit (%)
100
FiO2 Low Limit (%)
21
TVe High Limit (ml)
1000
TVe Low Limit (ml)
5
MV High Limit (L/min)
10
MV Low Limit (L/min)
2.0
Rate High Limit (BPM)
40
Rate Low Limit (BPM)
2
Paw High Limit (cmH
2
O) 30
Paw Low Limit (cmH
2
O) 4
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...