8.3.2.7 Set CO2 Waveform
1. Select the waveform area to access the waveform setup
menu.
2. Select
[
Waveform
] and select [
CO2
].
3. Select
[
Sweep
] and set waveform sweep speed to an appropriate value. The greater the
value is, the faster the waveform sweeps, the wider the waveform is.
4. Select
[
Scale
] and toggle between:
[
40
], [
60
] and [
80
] if the unit is mmHg;
[
5.0
], [
7.5
] and [
10.0
] if the unit is % or kPa.
5. Select to exit the current menu.
For details about displaying the CO2 waveform, refer to
5.3.3.2Display CO2 Waveform.
8.3.3 Measurement Limitations
Measurement accuracy may degrade due to:
Leakage or internal leakage of the sample gas.
Mechanical shock
Cyclic pressure which is greater than 10 kPa (100 cmH2O)
Other interference source (if available)
8.3.4 Troubleshooting
When the sampling system of the CO2 module works incorrectly, check if the sampling tube
is kinked. If not, remove the sampling tube from the watertrap. Then, if a prompt message
indicating airway malfunction appears on the screen, it means that the watertrap is occluded.
In this case, you must replace the watertrap. If no such prompt message is displayed, it means
that the sampling tube is occluded. Then you must replace the sampling tube.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...