13-17
CAUTION
z
Remember to do a breathing system leak test after reinstalling the sodalime
canister.
13.1.8.2 Change the Sodalime
NOTE
z
A gradual color change of the sodalime in the canister indicates absorption of
carbon dioxide. The color change of the sodalime is only a rough indicator. Use
carbon dioxide monitoring to determine when to change the sodalime.
z
Follow local regulations regarding disposal of hospital waste when the sodalime has
changed color. If left standing for several hours, it may regain its original color
giving a misleading indication of activity.
z
TM
Medisorb
sodalime is recommended.
1. Disassemble the sodalime canister by referring to
13.1.8.1Assemble the Sodalime
Canister
in the reverse order.
2. Pour out the sodalime which has changed color.
3. Press the canister support buckle to remove the canister support. Replace the sodalime
canister sponge.
Sodalime canister
sponge
Canister support
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...