13-5
3. Push the breathing system into the circuit adapter with force to let the breathing system
connected to the adapter seamlessly.
4. Set the locking catches on the circuit adapter to the
position and make sure that the
breathing system is safely locked.
WARNING
z
Set the locking catches to the
position after the breathing system is installed
onto the circuit adapter and make sure that the breathing system is reliably locked.
If not, the breathing system will be disconnected from the circuit adapter during
use, which can cause serious fresh gas leak and inaccurate tidal volume
measurement.