13-37
3. Connect the 30 mm female conical connector of the transfer tube to the AGSS waste gas
outlet on the anesthesia machine.
NOTE
z
Remove the AGSS transfer and receiving system from the main unit when
transporting or moving the anesthesia machine.
13.10.3 Waste Gas Disposal System
The AGSS transfer and receiving system is of high flow and low vacuum type, which is in
compliance with ISO 8835-3:1997. The applicable pump rate ranges from 50 to 80 L/min.
Before use, make sure that the waste gas disposal system is high-flow disposal system and is
able to reach the flow range.
Before use, make sure that the connector of the waste gas disposal system is BS6834-1987
standard connector.
For details about specifications, refer to
B.10AGSS Transfer and Receiving System
Specifications
.
NOTE
z
Do not block the pressure compensation opening of the AGSS transfer and
receiving system during test.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...