4-20
In PSV mode, you also need to set:
[
Trigger Level
]
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
Trigger Level
].
2. Select
[
Pressure
] or [
Flow
] for trigger type.
3. Turn the control knob to set [
Trigger Level
] to the appropriate value.
4. Push the control knob to confirm the setting.
5. Select
[
Ok
]
to activate the current setting.
6. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
[
PSV
Insp Termination Level
]
Inspiration termination level refers to the percentage of inspiration flow to the maximum
inspiration flow during inspiration in the PSV mode.
To set [
PSV
Insp Termination Level
], do as follows:
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
PSV
Insp Termination
Level
].
2. Push the control knob and turn it to set [
PSV Insp Termination Level
]
to the
appropriate value.
3. Push the control knob to confirm the setting.
4. Select
[
Ok
]
to activate the current setting.
5. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
[
Backup Mode Active
]
When PSV mode is applied alone, the PCV backup mode is available. If within the preset
time (Backup Mode Active), no spontaneous breathing occurs or the spontaneous breathing is
not strong enough to reach Trigger Level, the PCV backup mode is activated automatically
when the time for Backup Mode Active is up to enable mechanical ventilation forcibly.
To set [
Backup Mode Active
], do as follows:
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
Backup Mode Active
].
2. Push the control knob and turn it to set [
Backup Mode Active
]
to the appropriate value.
3. Push the control knob to confirm the setting.
4. Select
[
Ok
]
to activate the current setting.
5. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...