9-6
When [
Working
Mode
] is set to [
Measure
], the message [
AG Startup
] appears on the
screen. After start-up is finished, the message [
AG Warmup
] is displayed. The AG module is
in ISO accuracy mode .After warm-up is finished, the .AG module enters full accuracy mode.
9.5.5 Set CO2 Unit
In the [
Gas Module Setup >>
] menu, select [
CO2
Unit
] and toggle between [
mmHg
], [
%
],
and [
kPa
].
9.5.6 Restore Defaults
Select [
Defaults
] from the [
Gas Module Setup
] menu. Then all the options in this menu
except [
Working Mode
] are restored to the factory default configurations.
9.5.7 Set CO2 Waveform
1. Select the waveform area, open the corresponding
menu.
2. Select
[
CO2
] for [
Waveform
].
3. Select
[
Sweep
] and set waveform sweep speed to an appropriate value. The greater the
value is set to, the faster the waveform sweeps, the wider the waveform is.
4. Select
[
Scale
] to change the select CO
2
waveform scale. Options include:
[
40
], [
60
] and [
80
] if the unit is mmHg;
[
5.0
], [
7.5
] and [
10.0
] if the unit is % or kPa.
5. Select to exit the current menu.
For details about displaying the CO
2
waveform, refer to
5.3.3.2Display CO2 Waveform.
9.6 Change Anesthetic Agent
If the anesthetic agent used changes, the AG module is capable of detecting the gas mixture
during the transition period. The time required for anesthetic agent exchange depends upon
the type of anesthesia (low flow or high flow) and the features of the anesthetic agents used
(pharmacokinetics). During the exchange, the anesthesia machine gives no prompt message
and the MAC values displayed may be inaccurate.
The M-type AG module cannot identify anesthetic agent automatically. Therefore, you need
to change the setting of [
Agent
] to let the agent set consistent with the agent applied.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...