8-3
8.3 Use a Sidestream CO2 Module
NOTE
z
This section is only applicable to the anesthesia machine configured with
sidestream CO2 module.
8.3.1 Prepare to Measure CO2
1. Attach the watertrap to the watertrap fixer and then connect the CO2 components as
shown below.
2. By default, the CO2 module is in measure mode. The [
CO2 Startup
] message appears
on the screen when the CO2 module is plugged in.
3. After start-up is finished, the message [
CO2 Warmup
] is displayed. The CO
2
module is
in ISO accuracy mode. If you perform CO
2
measurements during warm-up, the
measurement accuracy may be compromised.
4. After warm-up is finished, the CO
2
module enters full accuracy mode.
NOTE
z
To extend the lifetime of the watertrap and CO2 module, disconnect the watertrap
and set the working mode of the module to standby when CO2 monitoring is not
required.
Watertrap
Watertrap fixer
Sampling tube
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...