B-20
Effect of interfering gas on AG measured value
Quantitive effect
(
%
ABS
)
2)
Gas
Concentration
(%)
CO2
N2O
Agent
O2
CO2 / /
0.1
0.1
0.2
N2O / 0.1
/
0.1
0.2
AG1) / 0.1
0.1
0.13)
1
Nitrogen
≤
78
%
0
0
0 0
Xenon
<
100% 0.1
0
0
0.5
Helium
<
50
%
0.1 0
0
0.5
Ethanol
<
0.1
%
0
0
0
0.5
Acetone
<
1
%
0.1 0.1 0 0.5
Methane
<
1
%
0.1 0.1 0 0.5
Methoxyflurane /
Unspecified
Unspecified
Unspecified
Unspecified
1) Multiple agent interference on CO2, N2O and O2 is typically the same as single agent
interference.
2) Maximum quantitive effect of each gas at concentrations within specified accuracy ranges for
each gas. The total effect of all interferences shall not exceed 5%REL of gas concentration.
3) Applicable to AION 03 AG module only, equivalent to the interference of secondary AG to
primary AG.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...