B
-6
Volume Approximately
6
ml
Interface and connector
Expiration end
Male 22 mm conical connector incorporating a coaxial female 15 mm
conical connector
Inspiration end
Male 22 mm conical connector incorporating a coaxial female 15 mm
conical connector
Bag end
Male 22 mm conical connector incorporating a coaxial female 15 mm
conical connector
Airway pressure gauge
Range
-20 to +100 cmH
2
O
Accuracy:
±2.5% of full range
APL valve
Range
1 to 75 cmH
2
O
Tactility indication
Above 30 cmH
2
O
Rotation range
1 to 30 cm H
2
O (0 to 145.8°)
30 to 75 cm H
2
O (145.8 to 292.5°)
Pressure flow data (APL valve completely open)
Flow (L/min)
APL pressure( cmH
2
O, dry)
APL pressure (cmH
2
O, moist)
3 0.23
0.24
10 0.25
0.25
20 0.27
0.27
30 0.28
0.28
40 0.30
0.31
50 0.33
0.34
60 0.36
0.40
70 0.41
0.46
Minimum pressure to open the APL valve
Dry 0.03
kPa
Moist 0.06
kPa
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...