13-28
13.4.3.2 Quik-Fil System
NOTE
z
To avoid spillage, check that the bottle to be used for draining has sufficient
capacity for the volume of liquid to be drained.
WARNING
z
The filler cap must be refitted before using the vaporizer.
1. Remove the protective cap from an empty bottle. Insert the bottle nozzle into the drain
funnel. Rotate the bottle to align the bottle filler keys with the index slots in the drain
funnel, and screw the drain funnel onto the empty bottle.
2. Remove the vaporizer filler block cap.
3. Fully insert the drain funnel into the keyed drain slot, and unscrew the drain plug.
Continue to drain the vaporizer until empty. Close the drain plug and tighten, and
withdraw the drain funnel.
4. Unscrew the drain funnel from the bottle and refit the bottle cap and the vaporizer filler
block cap.