10-6
10.7 Cyclic Impedance Check
This measures the exact impedance of each individual electrode. It causes a disturbed EEG
wave, and a prompt message is displayed on the screen.
The cyclic impedance check is automatically initiated when a sensor is connected. To
manually start a cyclic impedance check manually, you can either:
Select [
Cyc. Imped. Check
] in the [
BIS Module Setup
] menu and then select
[
ON
].
Press the
key on the BIS module.
Select [
Start Sensor Check
] in the BIS sensor check window.
The cyclic impedance check stops automatically if the impedances of all electrodes are
within the valid range. To manually stop a cyclic impedance check, you can either:
Select [
Cyc. Imped. Check
] in the [
BIS Module Setup
] menu and then select
[
OFF
].
Press the
key on the BIS module.
Select [
Stop Sensor Check
] in the BIS sensor check window.
10.8 BIS Sensor Check Window
To open the sensor check window, select [
Sensor Check >>
]
in the [
BIS Module Setup
]
menu. The graphic in the BIS sensor check window automatically adapts to show the type of
sensor you are using, show three or four electrodes as required. Each symbol in the graphic
represents an electrode and illustrates the most recently-measured impedance status of the
electrodes:
①
is the reference electrode;
②
the ground electrode;
③
and
④
are signal
electrodes
1
3 4
2
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...