B.11 O2 Sensor Specifications
O
2
sensor
Output
9-13 mV at 210 hPa O2
Range
0 to 1500 hPa O2
100% O2 signal deviation
100±1%
Resolution
1 hPa O2
Expected working life
1.5 x 106 % for measurement (20°C)
0.8 x 106 % for measurement (40°C)
Response time (21% air to 100% O2)
< 15 s
Linearity
Linear 0-100% O2
Operating temperature range
-20°C to +50°C
Temperature compensation
±2% of fluctuation at 0-40°C
Pressure range
50 to 200 KPa
Related humidity
0 to 99%
100% O2 concentration output drift
Over one year of typical value <5%
Material White
ABS
Packaging Sealed
package
Service life
Not more than 13 months after unpacked (in
compliance with the service conditions specified
by the manufacturer)
Effect of interfering gas
Gas under test
Error (% O2)
50% He/50% O2
<1%
80% N2 O/20% O2
1 to 1.5%
4% Halothane/28.8% O2 /67.2% N2O
1.5% to 2%
5% Sevoflurane/28.5% O2 / 66.5% N2O
1 to 1.5%
5% Enflurane/28.5% O2 /66.5% N2O 1.8% 1.2 to 1.8%
5% Isoflurane/28.5% O2 /66.5% N2O
1.2 to 1.8%
5% CO2 / 28.5% O2 /66.5% N2O
<1%
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...