7-4
5. Plug the Y piece into the leak test plug to close the breathing system.
6. Remove the water collection cup. For details, refer to
14.2.11 Water Collection Cup
.
Make sure that the system is Standby. If not, press the
key and then select [
Ok
] from the
pop-up menu to enter standby status.
Select the [
Maintenance
] shortcut key and then select [
Flow Sensor Cal. >>
] to open the
[
Flow Sensor Cal.
] menu. Select [
Start
] from the menu to start to calibrate the flow
sensor. The screen prompts [
Calibrating
].
9. During the calibration, if you select [
Stop
], calibration is stopped. Then the message
[
Calibration Stopped! Calibration is unfinished.
] is displayed. This indicates invalid
calibration instead of calibration failure.
10. After a successful calibration, the screen shows [
Calibration
Completed!
]. Otherwise,
the message [
Calibration Failure! Please try again.
] is displayed. In this case, you
need to do the calibration again.
11. Select to exit the current menu.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...