F-3
EtEnf Low Limit (%)
0.0
FiEnf High Limit (%)
2.0
FiEnf Low Limit (%)
0.0
EtIso High Limit (%)
3.0
EtIso Low Limit (%)
0.0
FiIso High Limit (%)
2.0
FiIso Low Limit (%)
0.0
EtSev High Limit (%)
6.0
EtSev Low Limit (%)
0.0
FiSev High Limit (%)
5.0
FiSev Low Limit (%)
0.0
EtDes High Limit (%)
8.0
EtDes Low Limit (%)
0.0
FiDes High Limit (%)
6.0
FiDes Low Limit (%)
0.0
F.3 BIS Module
BIS Setup
Factory default settings
Smoothing Rate
30 s
Cont. Imped. Check
ON
Cyc. Imped. Check
OFF
BIS module alarm limits
Alarm Level
Med
BIS High Limit
70
BIS Low Limit
20
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...