B-14
B.12 CO2 Module Specifications
Mainstream CO2 Module Specifications
Mainstream CO
2
module
Measurement mode
Mainstream
Measurement range
Accuracy
0 to 40 mmHg
±2 mmHg
41 to 70 mmHg
±5% of the reading
71 to 100 mmHg
±8% of the reading
Measurement range and
accuracy
101 to 150 mmHg
±10% of the reading
0 to 69 mmHg
0.1 mmHg
Resolution
70 to 150 mmHg
0.25 mmHg
Response time
<60 ms
Stability
Short-term drift: ±0.8 mmHg within 4 hours;
Long-term drift: accuracy specification retained within 120 hours.
Mainstream CO
2
module alarm specifications
CO
2
alarm limits
Range (mmHg)
Accuracy (mmHg)
Step (mmHg)
EtCO
2
High Limit
(low limit + 2) to 150
EtCO
2
Low Limit
0 to (high limit – 2)
FiCO2 High Limit
0 to 150
±1 1
Microstream CO2 Module Specifications
Microstream CO
2
module
Measurement mode
Microstream
Measurement range
Accuracy
0 to 38 mmHg
±2 mmHg
Measurement range
and accuracy
39 to 99 mmHg
±
5%
(
+0.08% for every 1mmHg above
38mmHg
)
Measurement
accuracy drift
Meets accuracy requirements within 6 hours
Resolution 1
mmHg
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...