14-15
14.2.13.3 Spoiler
After taking out the float, remove the spoiler.
14.3 Clean&Disinfect and Re-install the Breathing
System
Parts marked
are autoclavable. Metal and glass parts can be steam autoclaved.
Maximum recommended temperature is 134
º
C. By using autoclave to solidify
bacterioprotein rapidly, quick and reliable sterilization can be achieved. Suffered from 15 to
20 minutes of 1.05 kg/cm
2
steam pressure and 121ºC temperature, all bacteria and most brood
cells are killed.
Such parts are cleanable by hand. Rinse and dry all parts of the breathing system except the
O
2
sensor completely by using mild detergent (pH ranging from 7.0 to 10.5).
The flow sensor is plastic. For details about cleaning procedure, refer to
14.3.9 Flow Sensor.
WARNING
z
Do not use talc, zinc stearate, calcium carbonate, corn sarch or equivalent
materials to prevent tackiness. These materials can go into the patient’s lungs and
airways and cause irritation or injury.
z
Do not put both of the breathing system and the O2 sensor in liquid or autoclave
them.
z
Inspect all parts for deterioration. Replace them if necessary.
Spoiler
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...