13-12
3. Insert the other end of the O
2
sensor cable into the O
2
sensor connector marked
on
the circuit adapter.
13.1.8 Install the Sodalime Canister
WARNING
z
Obey applicable safety precautions.
z
Do not use the sodalime canister with chloroform or trichloroethylene.
z
Disposable sodalime canister is a sealed unit which should not be opened or
refilled.
z
Avoid skin or eye contact with the contents of the sodalime canister. In the event of
skin or eye contact, immediately rinse the affected area with water and seek
medical assistance.
z
Changing the sodalime during ventilation may result in breathing system leakage if
the anesthesia machine does not have BYPASS function.
z
If the anesthesia machine has BYPASS function, be sure to install and lock the
sodalime canister in place. Failure to do so will result in repeated inhalation of the
patient’s expired CO2.
z
CO2 concentration monitoring is strongly recommended when the anesthesia
machine has BYPASS function.
z
Before installing a sodalime canister, inspect the color of the sodalime in the
canister to determine when to change the sodalime.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...