4-3
4.3.2 Set Anesthetic Agent
NOTE
z
You do not need to perform this operation if inspiratory anesthetic agent is not
used.
z
This anesthesia machine can be mounted with vaporizers corresponding with
halothane, enflurane, isoflurane, sevoflurane and desflurane. Only one of the five
vaporizers can be opened at a time because the vaporizers are featured with
interlock.
4.3.2.1 Select the Desired Anesthetic Agent
1. Determine the anesthetic agent to be used and then fill the vaporizer. For details, refer to
13.4.2 Fill the Vaporizer.
2. Mount the vaporizer filled with anesthetic agent onto the anesthesia machine. For details,
refer to
13.4 Install the Vaporizer.
4.3.2.2 Adjust the Concentration of Anesthetic Agent
Push and turn the concentration control on the vaporizer to set the appropriate concentration
of anesthetic agent.
NOTE
z
Inspect the color of the sodalime in the canister before using the anesthetic agent.
Replace the sodalime immediately if obvious color change is detected.
z
For details about how to use the anesthetic agent, refer to the Vaporizer
Instructions for Use.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...