13-18
4. Pour new sodalime into the sodalime canister. When pouring, prevent the sodalime from
falling on the venthole of the canister support, which may increase airway resistance.
5. Install the canister support into the canister. Depress the canister support buckle to lock
the canister.
6. Assemble the sodalime canister.
WARNING
z
Do not reuse the sodalime canister sponge, which must be replaced every time the
sodalime canister is replaced.
z
The sodalime canister sponge must be in place to prevent dust and particles from
entering the breathing system.
z
When re-installing the sodalime canister after changing the sodalime, make sure
that the canister is locked reliably and installed in position.