The A-type AG module can identify anesthetic agent automatically. When one anesthetic
agent decreases below the threshold value and another anesthetic agent plays the dominant
role, the anesthesia machine can identify such exchange automatically and displays the name
and data of the dominant anesthetic agent.
9.7 Measurement Limitations
Measurement accuracy may degrade due to:
Leakage or internal leakage of the sample gas.
Mechanical shock
Cyclic pressure which is greater than 10 kPa (100 cmH2O)
Other interference source (if available)
9.8 Troubleshooting
If the gas inlet (including watertrap, sampling tube and airway adapter) is occluded by
condensed water, airway occlusion will be prompted on the screen.
To remove the occlusion:
Check the airway adapter for occlusion and replace if necessary.
Check the sampling tube for occlusion or kinking and replace if necessary.
Check the watertrap for water build-up. Empty the watertrap. If the problem persists,
replace the watertrap.
If the problem persists, internal occlusions may exist. Contact your service personnel.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...