D-13
BIS Overrange
H
SQI Overrange
H
SR Overrange
H
The measured value is outside of the measurement range.
Contact your service personnel.
BIS High Imped.
M
BIS Sensor Off
L
Check sensor connections. Re-connect the sensor.
BIS DSC Error
M BIS DSC signal reception error. Check the DSC.
BIS DSC Malf
H
BIS DSC switched off due to malfunction. Check the
DSC.
BIS No Cable
L
Connect the BIS cable.
BIS No Sensor
L
Connect the sensor.
BIS Wrong Sensor
Type
L
Check or replace the sensor.
SQI<50% L
SQI<15% L
SQI is too low. Check the patient’s condition and sensor
connection.
BIS Sensor Expired L
Replace the sensor.
BIS Sensor Failure
M Put the sensor again or replace the sensor.
BIS Sensor Too
Many Uses
L
Replace the sensor.
Disconnect/Reconn
ect BIS
H
Re-plug the BIS module.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...