D-11
MAC < 3
MAC value was less than 3.
Mixed Agent and
MAC >= 3
M More than more anesthetic agents were detected ant the
MAC value was not less than 3.
EtCO2 ALM LMT
ERR
H
The EtCO
2
alarm limit settings are outside of the range.
FiCO2 ALM LMT
ERR
H
The FiCO2 alarm limit settings are outside of the range.
CO2 Cal. Error
M An error occurred to CO
2
calibration.
CO2 Init Error
H
The CO
2
module was installed improperly or
malfunctioned.
CO2 Selftest Error
H
Module fault or communication failure between the
module and anesthesia machine. Re-plug the module,
restart the anesthesia machine, or try to plug the module
into another anesthesia machine.
CO2 Comm Stop
H
CO
2
module malfunction or communication failure
CO2 Comm Error
H
CO
2
module communication failure
CO2 Temp
Overrange
H
The temperature of the module crosses the range. Use the
module after it is kept away from the heat source or its
temperature falls within the normal range.
CO2 Sensor High
Temp
M The temperature of the sensor assembly is too high (>63
℃
). Check, stop using or replace the sensor.
CO2 Sensor Low
Temp
M The temperature of the sensor assembly is too low (<5
℃
).
Check, stop using or replace the sensor.
CO2 High Airway
Press.
M Paw is too high (>790 mmHg).An error occurred to the
airway pressure. Check the patient connection and
breathing system. Then restart the anesthesia machine.
CO2 Low Airway
Press.
M Paw is too low (<428 mmHg).An error occurred to the
airway pressure. Check the patient connection and
breathing system. Then restart the anesthesia machine.
CO2 High
Barometric
M The barometric pressure is greater than 790 mmHg. Check
the airway connections. Make sure that the anesthesia
machine application site meets the environmental
specifications. Check for special sources that affect the
ambient pressure. Restart the anesthesia machine.
CO
2
module
CO2 Low
Barometric
M The barometric pressure is less than 428 mmHg. Check
the airway connections. Make sure that the anesthesia
machine application site meets the environmental
specifications. Check for special sources that affect the
ambient pressure. Restart the anesthesia machine.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...