4-17
4.4.5.6 Parameter Range and Default Value in SIMV Mode
Parameter
Range
Step
Default
SIMV mode
TV
TV
20 to 1500 ml
20 to 100 ml: 5 ml
100 to 300 ml: 10 ml
300 to 1500 ml: 25 ml
SIMV-VC
Pinsp
5 to 60 cmH2O
1 cmH2O
15 cmH2O
SIMV-PC
SIMV Rate
4 to 60 BPM
1 BPM
10 BPM
Tinsp
0 to 4.5 s
0.1
1.5 s
Finsp
20 to 85 L/min
1 L/min
60 L/min
Plimit
10 to 100 cmH2O
1 cmH2O
30 cmH2O
Psupp
5 to 60 cmH2O
1 cmH2O
15 cmH2O
PEEP
OFF, 4 to 30
cmH2O
1 cmH2O
OFF
Trigger Window
5 to 90 %
5 %
25 %
Pressure
-20 to -1 cmH2O
1 cmH2O
-2 cmH2O
Trigger
Level
Flow
0.5 to 15 L/min
0.5 L/min
3 L/min
PSV Insp
Termination Level
5 to 60 %
5 %
25 %
SIMV-VC
SIMV-PC
4.4.6 Pressure Support Ventilation (PSV)
4.4.6.1 Description
Pressure support ventilation (hereinafter referred to as PCV) mode is an auxiliary breathing
mode which needs patient’s spontaneous breathing to trigger mechanical ventilation. When
the patient’s spontaneous inspiration reaches the preset Trigger Level, the ventilator begins to
deliver gas at the preset Finsp to make Paw rise to the preset Psupp rapidly. After that, the
ventilator slows down the flow through the feedback system to keep Paw constant. When the
inspiration flow drops to the preset PSV Insp Termination Level, the ventilator stops
delivering gas and the patient is allowed to expire, waiting for next inspiration trigger. If
inspiration is not triggered within the set time (Backup Mode Active), the system
automatically switches to the backup ventilation mode—PCV.
In the PSV mode, you do not need to set TV. TV depends on the patient’s inspiratory force
and pressure support level, compliance and resistance of the patient and of the whole system.
The PSV mode is used only when the patient is driven by reliable breathing because
breathing must be fully triggered by the patient during ventilation.
Summary of Contents for WATO EX-65
Page 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES ...
Page 16: ...8 FOR YOUR NOTES ...
Page 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Page 27: ...2 5 Breathing system ...
Page 29: ...2 7 2 2 2 Rear View Power supply ...
Page 31: ...2 9 Anesthesia information system CIS ...
Page 42: ...3 8 FOR YOUR NOTES ...
Page 66: ...4 24 FOR YOUR NOTES ...
Page 148: ...11 8 FOR YOUR NOTES ...
Page 152: ...12 4 FOR YOUR NOTES ...
Page 222: ...A 6 FOR YOUR NOTES ...
Page 244: ...B 22 FOR YOUR NOTES ...
Page 250: ...C 6 FOR YOUR NOTES ...
Page 264: ...D 14 FOR YOUR NOTES ...
Page 274: ......
Page 275: ......
Page 276: ...P N 046 000203 00 1 0 ...